-40%

National Instruments SCXI-1120 8-Channel Isolation Amplifier Input Module Card

$ 395.99

Availability: 100 in stock
  • Condition: These units are used, old stock.
  • Brand: National Instruments
  • All returns accepted: ReturnsNotAccepted
  • MPN: SCXI-1120

    Description

    National Instruments SCXI-1120 8-Channel Analog Isolation Amplifier Input Module
    Comes with a manul and what you see in the pictures. If you don't see it, you probably wont get it.
    Specifications are from National Instruments and may vary slightly due to upgrades, options, or revisions this unit may or may not have.
    These units are used, old stock.
    The Unit's Serial Number Tag Reads:
    Model Number:
    SCXI-1120
    Part Number:
    181695J-01
    Revision:
    001
    Serial Number:
    C52286
    CE Marked:
    Yes
    Key Features:
    8 channels
    333 kS/s maximum sampling rate
    Gain and lowpass filter settings per channel
    Signal inputs from ±2.5 mV to ±1000 VDC with TBX-1316
    Jumper selectable filter per channel - 4 Hz and 10 kHz filter
    250 V
    rms
    working isolation per channel
    Description:
    The National Instruments SCXI-1125, SCXI-1120, and SCXI-1120D are 8-channel isolated analog input modules. These modules share a common architecture, providing 250 to 300 V
    rms
    of working isolation and lowpass filtering for each analog input channel. This  architecture is ideal for amplification and isolation of millivolt, volt, 0 to 20 mA, 4 to 20 mA, and thermocouple signals. Each module can multiplex these eight channels into a single  channel of the DAQ device, and you can add modules to increase channel count. These modules also offer parallel mode operation for increased scanning rates.
    The analog inputs of the NI-1120/D consist of eight isolation amplifiers. You can configure each amplifier using jumpers for input ranges from ±2.5 mV to ±5 V. With the SCXI-1327 high-voltage attenuator terminal block, the input range is extended to ±250 V. With the TBX-1316, the input range is extended to ±1000 VDC (680 V
    rms
    ). Each channel also includes a lowpass filter that is jumper configurable for 4 Hz or 10 kHz. Each channel is individually isolated with a working common-mode voltage of 250 V
    rms
    between channels or channel to eart. In addition the SCXI-1120 is CE certified as double insulated, Category II, for 250 V
    rms
    of operational isolation.
    Analog Input Specifications:
    Gain (Jumper-selectable):
    1, 2, 5, 10, 20, 50, 100, 200, 250, 500, 1,000, 2,000
    Output Range:
    ±5 V
    Number of Channels:
    8
    Gain Accuracy:
    0.2% of reading typ., 0.6% of reading max.
    Nonlinearity:
    0.04% of fullscale range
    Offset Voltage:
    ±[6 µV + (3 mV / Gain)]
    Stability Verses Ambient Temperature:
    Offset Drift RTI:
    ±[0.2 µV/°C ± (200 µV/°C / Gain)]
    Gain Drift:
    20 ppm/°C
    Input Bias Current:
    ±80 pA
    Input Resistance:
    Normal:
    1 GΩ
    Power off:
    50 kΩ
    Overload:
    50 kΩ
    Output Resistance:
    Multiplexed-output Mode:
    100 Ω
    Parallel-output Mode:
    330 Ω
    Filtering (Jumper Selectable):
    4 Hz (-10 dB) or 10 kHz (-3 dB), 3-polse RC
    Output Selection Time (with 5 V step, all gains):
    0.012% Accuracy
    1
    :
    5.2 µs typ., 7 µs max.
    0.012% Accuracy
    2
    :
    3 µs
    0.006% Accuracy
    3
    :
    10 µs
    0.0015% Accuracy
    3
    :
    20 µs
    Rise Time:
    4 Hz:
    0.12 s
    10 kHz:
    70 µs
    Slew Rate:
    0.15 V/µs
    Nose (400 kHz Bandwidth)
    1
    :
    Input (Gain = 1,000):
    4 Hz Filter:
    100 nV
    rms
    10 Hz Filter:
    4 µV
    rms
    Output (Gain = 1):
    4 Hz Filter:
    150 µV
    rms
    10 kHz Filter:
    1 mV
    rms
    Operating Common-Mode Voltage:
    50 or 60 Hz
    Channel to Channel or Channel to Earth:
    250 V
    rms
    Common-Mode Rejection Ratio, 50 or 60 Hz:
    1 kΩ in Input Leads:
    160 dB min at 4 Hz bandwidth
    NMR (50 or 60 Hz):
    60 dB at 4 Hz bandwidth
    Input Protection (Continuous):
    250 V
    rms
    max
    1
    Output Protection:
    Continuous short-to-ground
    Power Consumption:
    170 mA at ±18.5 V, 50 mA at ±5 V
    Notes:
    1:
    Includes the combined effects of the SCXI-1120 and the AT-MIO-16F-5
    2:
    Includes the combined effects of the SCXI-1120 and the AT-MIO-16E-2 with 1 m or 2 m cable
    3:
    Includes the combined effects of the SCXI-1120 and the AT-MIO-16X with 1 m or 2 m cable
    Digital I/O Specifications:
    Absolute Maximum Voltage Input Rating:
    5.5 V with respect to DIG END
    Digital Input Specifications (Referenced to DIG GND):
    V
    IH
    Input Logic High Voltage:
    2 V min
    V
    IL
    Input Logic Low Voltage:
    0.8 V max
    I
    I
    Input Current Leakage:
    ±1 µA max
    Digital Output Specifications (Referenced to DIG GND):
    V
    OH
    :
    3.7 V min at 4 mA max
    V
    OL
    :
    0.4 V max at 4 mA max
    Physical Specifications:
    Dimensions:
    3.1 cm x 17.3 cm x 20.3 cm (1.2" x 6.8" x 8.0")
    Connectors:
    50-pin male ribbon-cable rear connector, 32-pin DIN C male front connector (18-screw terminal adapter available from National Instruments)
    Environment Specifications:
    Operating Temperature:
    0 to 50°C
    Storage Temperature:
    -20 to 70°C
    Relative Humidity:
    5% to 90% non-condensing
    Safety:
    Designed in accordance with IEC 61010-1, UL3111-1, and CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment. Approved at altitudes up to 2000 m, Installation Category II, Pollution Degree 2
    Notes:
    1:
    Module designed to meet IEC-1010 standard for double or reinforced insulation of category II pollution degree 2
    For Domestic Customers: Packaging, handling, and order processing included in shipping in all domestic shipments as quoted by the shipping calculator. For International Customers there is a .00 minimum for packaging, handling, and order processing. International shipping to be determined by destination.
    Prior to shipment I will need to know the following information (filled out on a form that I will send you) for International Customers or Packages/Freight being Forwarded internationally. International customers may also be subject to a freight forwarder form if applicable.:
    (1) Are you the end-user of this item?
    (2) If you are not the ultimate end-user of the item, please state the ultimate end user's name.
    (3) What is the ultimate country destination?
    03/13/15